CVE-2025-4395

MEDIUM
6.8CVSS v3
CVSS v2
0.27% EPSS (exploit probability)
CWE-258CWE

Description

Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access modify system functionality.

This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025

CVSS v3 vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Affected routers (0)

No routers currently mapped to this CVE in our database.

External references