CVE-2025-15653
MEDIUM6.8CVSS v3
—CVSS v2
0.02%
EPSS (exploit probability)
CWE-668CWE
Description
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
CVSS v3 vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Affected routers (0)
No routers currently mapped to this CVE in our database.